Abstract

The diagnosis of periprosthetic joint infection (PJI) remains a labor-intensive and challenging issue, with life-threatening complications associated with misdiagnoses. Superior diagnostic approaches are therefore urgently needed, and synovial biomarkers are gaining substantial attention in this capacity. A new aptamer-based sandwich assay was developed where the aptamer probes specific to one such biomarker, alpha-defensin human neutrophil protein 1 (HNP 1), was integrated herein into a new microfluidic platform. The magnetic beads coated with the primary aptamer probe were able to bind the target protein with high affinity and high specificity in synovial fluid and a fluorescent-labelled secondary aptamer were further used to quantify HNP 1 in a sandwich approach. Up to four clinical samples with low volume (∼50μL each) in a much faster assay including detection within <60min with 100% accuracy (with totally 13 clinical samples without the need of sample pretreatment) through the use of the aptamer-based sandwich assay were automatically detected on a single chip. The wide dynamic range of this compact device, 0.5-100mg/L, highlights its utility for future PJI diagnostics in the clinic.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.