Abstract

Thin-walled vacuum containment vessels cooled by circulating water jackets are often utilized in research and industrial applications where isolation of equipment or experiments from the influences of the surrounding environment is desirable. The development of leaks in these vessels can result in costly downtime for the facility. A Vessel Integrity Monitoring System (VIMS) is developed to detect leak formation and estimate the size of the leak to allow evaluation of the risk associated with continued operation.A wide range of leak configurations and fluid flow phenomena are considered in the evaluation of the rate at which a tracer gas dissolved in the cooling jacket water is transported into the vacuum vessel. A methodology is presented that uses basic fluid flow models and careful evaluation of their ranges of applicability to provide a conservative estimate of the transport rates for the tracer gas and hence the time required for the VIMS to detect a leak of a given size.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.