Abstract

Copper (Cu), a natural micronutrient with ecotoxicological significance, is involved in the carbon and nitrogen cycles occurring in marine ecosystems. Here, we developed a novel, antifouling gel-protected iridium (Ir) needle electrode modified with gold nanoparticles (G-IrNS) for long-term continuous and steady Cu monitoring. The gel formed an efficient membrane that effectively prevented the fouling of the sensing surface and displayed anti-convective properties, ensuring that mass transport toward the sensor surface was wholly controlled via diffusion. The repeatability, reproducibility, and stability of G-IrNS showed that it was suitable for long-term and on-site monitoring of Cu in seawater. Cu concentrations were successfully measured via fixed-point continuous monitoring for >2 weeks and onboard continuous monitoring in Bohai Sea using one sensor. Moreover, the relationship between Cu concentrations measured on-site via G-IrNS and its dissolved concentration in Bohai Sea was evaluated. G-IrNS can be applied to other metal ions as well, especially for long-term automatic on-site monitoring, thereby providing a basis for further research.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call