Abstract

An analytical formula for the electrical impedance between an adherent living cell and a sensor substrate measured using a microelectrode is presented for the first time. Previously-reported formula has been applicable only for the case where many cells are on a large electrode. In contrast, our formula is valid even when a microelectrode smaller than the cell-size is underneath the cell, which is often the case for the state-of-the-art single-cell analysis. Numerical simulations for verifying the accuracy of our formula reveals that the discrepancies between the theoretical impedances calculated by our formula and numerical simulation results are negligibly small. Our formula will be useful for describing cell-substrate impedance properties in equivalent circuit model analysis or sensor design optimizations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.