Abstract
A novel glucose biosensor was fabricated by immobilizing glucose oxidase (GOD) on a platinum electrode modified with NiO doped ZnO nanorods (NRs), which were adopted to construct an amperometric glucose biosensor for the first time. ZnO single-crystalline nanorod and NiO doped ZnO nanorods are prepared by hydrothermal method and characterized by X-ray diffraction (XRD) and scanning electron microscope (SEM). The electrochemical behavior of the platinum electrode modified with NiO doped ZnO nanorod was investigated in phosphate buffer solution (PBS). The effect of molar ratio of Ni/Zn in the doped ZnO nanorod, pH value, applied potential, temperature was investigated in detail. Under the optimal experimental condition, the constructed glucose biosensor showed a good linearly relationship between the concentration of glucose and the electrochemical current, a high sensitivity of 61.78μA/mMcm2, and a fast response time of <5s. All results demonstrate that the NiO doped ZnO NRs can provide a new platform for glucose biosensor design and other biological applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.