Abstract

This paper presents a cuff-less hypertension pre-screening device that non-invasively monitors the Blood Pressure (BP) and Heart Rate (HR) continuously. The proposed device simultaneously records two clinically significant and highly correlated biomedical signals, viz., Electrocardiogram (ECG) and Photoplethysmogram (PPG). The device provides a common data acquisition platform that can interface with PC/laptop, Smart phone/tablet and Raspberry-pi etc. The hardware stores and processes the recorded ECG and PPG in order to extract the real-time BP and HR using kernel regression approach. The BP and HR estimation error is measured in terms of normalized mean square error, Error Standard Deviation (ESD) and Mean Absolute Error (MAE), with respect to a clinically proven digital BP monitor (OMRON HBP1300). The computed error falls under the maximum standard allowable error mentioned by Association for the Advancement of Medical Instrumentation; MAE <; 5 mmHg and ESD <; 8mmHg. The results are validated using two-tailed dependent sample t-test also. The proposed device is a portable low-cost home and clinic bases solution for continuous health monitoring.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.