Abstract

This paper addresses the development of an aortic pressure observer for the Penn State Electric Ventricular Assist Device (EVAD). The observer estimates the aortic blood pressure by measuring the voltage of the electric motor and the pusher plate position. The estimated pressure is fedback to the EVAD's blood flow controller, which adjusts the beat rate of the device to accommodate the varying demand of cardiac output. The gains of the observer are deterministically optimized such that the optimal values are independent of the (often unknown) system state initial conditions. To improve the performance of the pressure observer an adaptation scheme is developed. In this scheme the initial pressure estimate of the succeeding systolic cycle is adjusted when the pressure does not match its corresponding second estimated value. In vitro test runs of the developed observer show that is is robust to parameter variations, and the error of the resultant estimated pressure is less than 5%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.