Abstract
An output feedback controller for the Penn State rollerscrew electric ventricular assist device (EVAD) is discussed. The controller minimizes the electrical energy consumption while using the only observable state variable as its signal. The initial conditions on which the optimal control of the EVAD depends are based on cardiovascular data of the general population. Comparison of the Rollerscrew EVAD with the Cam EVAD reveals a striking difference in efficiency between the two devices. A patient with an optimally controlled Rollerscrew EVAD could function for 14.5 h without a recharge while the same patient with an optimally controlled Cam EVAD could function only for 8 h. >
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.