Abstract
A flow injection analysis (FIA) method, based on the molybdenum blue formation, has been adapted and successfully optimized in order to use it to measure phosphate in estuarine water samples. Matrix effects have been avoided using the highest peak, and the Schlieren effect (common in FIA analysis) has been minimized by injecting the ascorbic acid with the same sulphuric acid concentration as the molybdate reagent. The concentrations of the reagents have been optimized by a Central Composite Design. The measuring range used was 0.02–0.2mgL−1 phosphate and both univariate and multivariate regressions have been compared using a random sample, reaching the same concentration (0.065±0.002mgL−1). However, the univariate regression showed lower LODs, 7μgL−1 (74 nM) in front of the 10μgL−1 (105 nM) reached by the multivariate one. The effectiveness of the developed method was tested with real samples. For this, low-cost conventional equipment easily available to most laboratories has been used, with a LOD low enough to measure estuary samples, and high repeatability.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.