Abstract

In the Electrotechnical Laboratory (ETL) the accuracy of the AC-DC voltage transfer difference was improved for the voltage range of 1-20 V in the frequency range from 10 Hz to 100 kHz. At the same time, the AC-Dc transfer standards and the measurement system which is used in the Japan Electric Meter Inspection Corporation (JEMIC) were also improved. The system is realized by applying an improved differential and a dual-channel method. The test repeatability is better than 1 ppm for nominal test voltages. The voltage and frequency measurement ranges are extended up to 200 V and up to 1 MHz. The whole system is controlled by a desktop computer with an optically isolated GP-IB bus system. >

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.