Abstract
An intercomparison of calibration systems for AC shunts up to audio frequencies (10 kHz) between the National Research Council of Canada, Japan Electric Meters Inspection Corporation, and the National Institute of Standards and Technology, is presented. The comparison was implemented with a transfer standard of 10 A, 0.1 /spl Omega/ calculable AC/DC shunt, designed by Japan Electric Meters Inspection Corporation. The results indicate that there are no significant differences in the overall accuracy of calibration systems for AC shunts at frequencies up to 10 kHz in each laboratory.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.