Abstract

Cascaded Talbot-Lau interferometers (CTLI) facilitates X-ray differential phase-contrast imaging (DPCI) by using absorption gratings with large periods. Large-area gratings provide a large field of view (FoV), thus they extend applications of DPCI. However, conventional methods for fabricating such gratings are still costly. We report an 8 inch absorption grating that was fabricated by filling tungsten nanoparticles into grating structures. The fabrication process requires readily available equipment, allowing it to be easily implemented. We experimentally tested these absorption gratings with our X-ray CTLI system, yielding Moiré fringes and multi-contrast images, which demonstrates the effectiveness of our proposed method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.