Abstract

A novel catechol biosensor was described based on the immobilization of polyphenol oxidase (PPO) into polyaniline (PANI), which was easily constructed by direct electropolymerization of aniline in a solution containing ionic liquid, 1-ethyl-3-methylimidazolium ethyl sulfate (EMIES). The developed biosensor for the detection of catechol has a linear range of 1.25–150μmoldm−3. The maximum response current (Imax) and the Michaelis–Menten constant (k′m) are 0.62μA and 146μmoldm−3, respectively. The activation energy (Ea) of the PPO catalytic reaction is 31.1kJmol−1 in the B–R buffer. The biosensor shows good reproducibility (a relative standard deviation of 3.1% was obtained) and remarkable long-term stability (it retains 75% of the original activity after four months). The effects of potential and pH on the response current of the biosensor are also described.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.