Abstract
As a research project for aluminum alloy molecular beam epitaxy systems, a large aluminum alloy chamber with a special surface finish was constructed. Its outgassing rate and ultimate pressure were measured. The special surface finish was analyzed using Auger electron spectroscopy. The oxide layer on the aluminum alloy was about 40 A thick. The outgassing rate before baking was measured as 10−11 Torr l/s cm2, and the rate after baking (150 °C, 24 h) was 3×10−13 Torr l/s cm2. This value is approximately one to two orders of magnitude lower than that of a stainless‐steel chamber, such as SUS 304. The ultimate pressure was 7×10−11 Torr when pumped with a turbomolecular pump and an ion pump. With addition of the newly developed titanium sublimation pump, the ultimate pressure was 6×10−12 Torr (6.5×10−12 mbar). This ultimate pressure is very low for a large aluminum alloy vacuum chamber. Ceramic coating of SiO2 using a spark discharge in a silicate solution on aluminum alloy was not corroded by gallium. This c...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Vacuum Science & Technology A: Vacuum, Surfaces, and Films
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.