Abstract

Other researchers have proposed sixport reflectometer calibration procedures that require only four standards. The memorandum describes an alternative approach to the implementation of such a calibration routine. The standards required are a matched load, a short circuit and two offset terminations. Laboratory tests indicate that measurement accuracies of within ±0.01 in magnitude and within ±1° in phase can be obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.