Abstract

Frequent monitoring of serum alanine aminotransferase (ALT) activity is essential to prevent drug-induced liver injury (DILI). Current ALT assays are restricted to centralized clinical laboratories, making frequent patient monitoring logistically difficult. To address this, we demonstrated the capability of commercial foundry manufactured silicon nanowire field effect transistor (SiNW-FET) biosensors in a form factor that enables frequent near-patient monitoring. Here, we designed an ALT assay, by coupling the ALT-catalyzed production of pyruvate to the reduction of ferricyanide, enabling both spectrophotometric and electrical measurement of ALT activity. The two methods yield comparable ALT activity detection across a dynamic range wide enough to monitor patients at risk for DILI. This study demonstrates kinetic activity measurement of an endogenous enzyme using uncoupled SiNW-FETs, and commercial manufacturing of SiNW-FET sensor arrays for use in a portable biosensor platform.