Abstract

The goal of this study was to analyze the agreement between the Medmont E300 and the Visionix VX120+ systems in terms of non-invasive tear break-up time (NIBUT) and tear meniscus height (TMH) measurements. A total of 60 eyes (30 healthy subjects) were enrolled. NIBUT and TMH were evaluated with Medmont E300; first NIBUT, NIBUT50%, and TMH were evaluated with Visionix VX120+. Both evaluations were performed in a random order by the same clinician for right, left, and both eyes. The Medmont E300 provided significantly higher NIBUT than Visionix VX120+ for first NIBUT in right, left, and both eyes (p ≤ 0.003) and NIBUT50% in left and both eyes (p ≤ 0.042). The TMH measured with VX120+ was significantly higher than with Medmont E300 considering both eyes (p = 0.037). No significant correlations were found between both devices for either NIBUT (p ≥ 0.11) or TMH (p ≥ 0.09). Passing–Bablok regression analyses revealed poor agreement between devices for NIBUT and TMH outcomes. VX120+ is expected to provide substantial lower first NIBUT values than the NIBUT measured by Medmont E300. Clinicians should consider not using both instruments as interchangeable for dry eye diagnosis.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call