Abstract
This study presents the results of an aging stability study of complete blood count (CBC) and leukocyte differential parameters using the Abbott CELL-DYN Sapphire hematology analyzer. Stability studies showed no substantial change in CBC parameters up to 24-48 h at +23 +/- 2 degrees C (room temperature), except for optical platelet count (PLTo). For specimens aged over 24, the value of impedance platelet count yielded more reliable results than the routine PLTo. White blood cell (WBC) differential parameters, except eosinophils, were stable for up to 48 h at +23 +/- 2 degrees C. CBC parameters were stable for 72 h, except mean platelet volume, which slightly increased between 48 and 72 h, at +4 degrees C. WBC differentials were stable 48-72 h, with a slight decrease observed in absolute neutrophils and lymphocytes at +4 degrees C.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.