Abstract

We present the results of the tests performed on 90 photomultiplier tubes (PMT) to characterize their afterpulses. Three different types of Hamamatsu PMTs (R7525, R6427, and R1398) were studied for their afterpulse rates and timings at different incident light intensities and gain values, at the University of Iowa PMT test station. Afterpulse rates show slight increase with the PMT gain, but there is almost no dependence on incident light intensity. Three specific time delays are determined for the afterpulses, and their individual rate contributions are characterized. The results from manufacturer's independent tests on R7525 PMTs are reported, as well. The possible effects of these afterpulses on the future hadron collider experiments are also discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.