Abstract
Phototimer set-up is a critical procedure for dose and image quality optimisation in computed radiography (CR) systems. While a conventional radiography automatic exposure control device (AEC) can be calibrated in order to gain a constant optical density on the film independent of beam quality and patient size, CR detectors present a high dynamic range which allows a much larger dose interval, but with different image quality levels. CR leads to a less frequent exam repetition, but may produce quite noisy images if the exposure level on the plate is not correct. The aim of this work is to evaluate the performance of a CR plate (Agfa MD40) in order to optimally calibrate an AEC device. The plate response has been characterised in terms of digital signal, exposure on the plate and signal-to-noise ratio for different beam qualities, in a patient of standard size.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.