Abstract

Recent clinical failures and abnormal sounds observed in certain prosthetic hip designs have directed attention to adverse condition hip simulator testing. In the present study, hip simulator wear and friction tests were made with a macroscopic separation of the bearing surfaces in the swing phase, steep acetabular cup position, increased load, poor or lacking lubrication, roughened bearing surfaces including titanium third bodies, and with combinations of these. The only conditions that resulted in squeaking were dry sliding with alumina-on-alumina and metal-on-metal, and an extreme peak load of 4kN with serum lubricated metal-on-metal. The wear rate of metal-on-metal increased by an order of magnitude when the cup position was steep, 64°, compared with the optimal position of 48°. The increase of the peak load from 2kN to 3kN with the cup position of 48° increased the running-in wear of metal-on-metal, but the steady state wear rates were equally low, of the order of 1mg per one million cycles. Crosslinked polyethylene was the superior cup material under adverse conditions, including dry sliding and roughened femoral head.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.