Abstract

The application of adsorptive stripping potentiometry to the reductive detection of nucleic acids at mercury electrodes is reported. Compared to analogous voltammetric stripping modes, constant current potentiometric stripping analysis (PSA) effectively addresses the hydrogen discharge background problem, and hence greatly improves the characteristics of the superimposed cytosine/adenine (CA) reduction peak. Compared to earlier schemes for trace measurements of nucleic acids at mercury or carbon electrodes that rely on anodic signals arising from the guanine residue, convenient quantitation can now be carried out in connection with the cytosine and adenine residues. Variables influencing the adsorptive PSA response are explored and optimized. With five minute accumulation, the detection limits for tRNA, ssDNA and dsDNA are 30 μg 1 −1, 60 μg 1 −1 and 2 mg 1 −1, respectively. Such different values reflect the strong dependence of the PSA CA signal upon the nucleic-acid structure. This allows the quantitation of ssDNA or tRNA in the presence of dsDNA, and offers new possibilities for electrochemical studies of DNA structure and interactions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.