Abstract

The adsorption of cytochrome c (cyt c) to a silica surface has been studied by use of evanescent wave broadband cavityenhanced absorption spectroscopy (EW-BBCEAS). Visible radiation from a supercontinuum source is coupled into an optical cavity consisting of a pair of broadband high reflectivity mirrors, and a total internal reflection (TIR) event at the prism/water interface. Aqueous solutions of cyt c are placed onto the TIR footprint on the prism surface and the subsequent protein adsorption is probed by the resulting evanescent wave. The time integrated cavity output is directed into a spectrometer, where it is dispersed and analysed. The high spectral brilliance of the SC affords a baseline noise comparable to evanescent wave cavity ring-down spectroscopy (EW-CRDS), and the broadband nature of the source allows observation of a wide spectral range (ca 250 nm in the visible). The system is calibrated by measuring the absorption spectra of dyes of a known absorbance. Absorption spectra of cyt c are obtained for both S and P polarized radiation, allowing information about the orientation of the adsorbed protein to be extracted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.