Abstract

Low light level (LLL) calibration becomes more and more important since the rapid growth of remote sensing. The spectral radiance at normal higher light levels can be calibrated with good accuracy, while LLL spectral radiance cannot. If an adjustable light source can be designed at nearly constant correlated color temperature (CCT) covering several orders of magnitude, low light level spectral radiance can be obtained with the help of a photodetector. Whether or not the spectral distribution of an integrating sphere based light source is nearly constant is investigated. By adjusting the diameter of the variable aperture between the integrating sphere and tungsten lamp, the spectral radiance can be varied over 6 orders of magnitude. However, the relative spectrum in the red region increases notably when the spectral radiance is decreased to 1/100000. If the spectral radiance is decreased further, the spectral difference can be more than 300% and CCT decreases more than 250 K. By using baffles and another integrating sphere, low light level radiation source at nearly constant spectral distribution is obtained. The variation of CCT is less than 50 K over 6 orders of magnitude.