Abstract

The spectral radiance measurement at daytime level can be realized with high accuracy, while it’s difficult when the spectral radiance is at nighttime level. We design a spectral radiance calibration facility which has the characteristics of completely unchanged spectrum over 3 orders of magnitude and approximately unchanged spectrum for about 6 orders of magnitude. It combines a spectral radiance light source, a precision aperture and a white diffuser together, make it easy to reproduce the spectral radiance at 380 nm from 4 × 10−9 W/(m2·sr·nm) to 4 × 10−3 W/(m2·sr·nm). The facility can be easily used to calibrate a spectroradiometer at nighttime level. When the spectral radiance from 380 nm to 780 nm is around 1 × 10−7W/(m2·sr·nm), the calibration uncertainty of the spectroradiometer is 0.87%∼1.0% (k = 1).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.