Abstract

The purpose of this study was to assess the early changes in stenosis geometry after insertion of intravascular stents in human coronary arteries. Morphologic changes were evaluated by quantitative coronary angiography (using automated edge detection) and by calculation of the theoretical pressure decrease across the dilated and stented stenosis from the Poiseuille and turbulent resistances assuming a coronary blood flow of either 1 or 3 ml/s. Twenty-six patients were studied before and after angioplasty, as well as immediately after stent implantation. The stented coronary artery was the left anterior descending artery in 19 cases, the circumflex artery in 2 cases, the right coronary artery in 2 cases and a coronary artery bypass vein graft in 3 cases. After stent implantation, an additional increase in minimal luminal cross-sectional area of the dilated vessel was observed, suggesting that the self-expanding stainless steel endoprosthesis used in this study has a dilating function in addition to its stenting role.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.