The objective of this study was to assess the efficacy of adaptive signal enhancement (ASE) as a means of indicating intraoperative spinal cord impingement. ASE technique was used to determine the changes in the somatosensory evoked potential (SEP) elicited from eighteen rats with varying levels of spinal cord compression. ASE technique was found to be able to effectively extract SEP signals for the detection of spinal cord injury. Furthermore, while the traditional ensemble averaging (EA) technique requires more than 500 trials for meaningful signal processing in severe noisy SEP recordings, the ASE method required only 50 trials to provide similar information. Because of its fast and reliable SEP detection, the ASE method is ideal for spinal cord monitoring in the clinical setting.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.