Abstract

Electrogastric measurements are useful for medical research and in clinical diagnosis. The measurements, however, contain very heavy respiratory artefact. Existing conventional frequency-domain filters cannot be used because of the possible overlap of the frequencies of the gastric signals and respiratory artefact. In the paper, the methods of measuring cutaneous and intraluminal gastric signals and reference respiratory signals are described. An adaptive cancellation technique is developed, which is simple and easy to implement for online processing. It is proved by experiments to be very efficient, i.e. the respiratory artefact can be completely cancelled, while the gastric signal component is not affected. Other possible applications in biomedical signal analysis are also discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.