Abstract

A second generation prototype enabling surface plasmon resonance spectroscopic measurements in the infrared (IR) range is described. The new design (v2) uses the optical train (optics and detector) within conventional FT-IR spectrometers by confining dimensions of the accessory to space available within the sample compartment of the spectrometer. The v2 accessory builds upon knowledge gained from a previous version that was based on a modified commercial variable angle spectroscopic accessory and addresses observed limitations of the original design-improved temporal stability and measurement acquisition speed, crucial to biomolecular binding studies, as well as optical flexibility, a requirement for investigations of novel plasmon-supporting materials. Different aspects of the accessory, including temporal stability, mechanical resilience, and sensitivity to changes in refractive index of a sample were evaluated and presented in this contribution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.