Abstract
A 32-year-old woman underwent transcatheter closure of a secundum type atrial septal defect with the Amplatzer device. The procedure was started under premedication with aspirin, clopidogrel, and heparin. During the procedure, a highly mobile thrombus attached to the left atrial disc of the device was detected by transesophageal echocardiography (TEE). The device and the associated thrombus were successfully withdrawn and the patient was started on a combination of heparin and tirofiban infusion. The procedure was successfully completed without any recurrent thrombus formation or residual shunt. Further investigation for thrombophilia revealed homozygous factor V Leiden mutation and the patient was started on a life-long warfarin therapy. Follow-up TEE showed proper device position with no recurrent thrombus and the follow-up was uneventful.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Turk Kardiyoloji Dernegi Arsivi-Archives of the Turkish Society of Cardiology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.