Abstract

A 49-year-old man had uneventful endocapsular phacoemulsification with in-the-bag implantation of an AcrySof ® SA60AT single-piece intraocular lens (IOL) (Alcon) in the right eye. Twenty-seven days postoperatively, he presented with ocular pain, intraocular pressure of 48 mm Hg, 360 degrees of hyperpigmentation of the trabecular meshwork, and iris pigment epithelial atrophy in the region of the upper temporal haptic, which had dislocated into the sulcus. The patient made an excellent recovery following IOL removal and exchange. Scanning electron microscopy of the explanted IOL demonstrated that the haptic had a rough lateral surface and anterolateral edge. We do not think this IOL should be implanted in the sulcus

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.