Abstract
An approximate solution for the acoustic coupling factor (the diffraction correction function) from a focused transducer to a flat plate and back to the transducer is provided. This function is useful for system calibrations where a pulse-echo system or transmit-receive system is used. Numerical solutions are provided for the important case where the flat plate is placed near the focal plane of the transducer. The solution for a flat disk transducer is obtained as a limiting case. Experimental evidence for a focused transducer is provided.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.