Abstract

Multicollection system allows improvement of measurement reproducibility for isotope ratios on SIMS instruments provided that the other sources of measurement variability are well controlled. This paper presents analytical protocols developed to control the mass fractionation in the collection optical system and the variation of electron multiplier yield. Results for both silicon and oxygen isotope ratios demonstrate that an experimental error as low 0.2 per mil is achievable whenever both isotopes were measured either with Faraday cup detectors or combining one Faraday cup and one electron multiplier (EM).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.