Abstract

The design, operating parameters and the accuracy evaluation of the NPL Rb atomic fountain are described. The atomic fountain employs a double magneto-optical arrangement that allows a large number of 87Rb atoms to be trapped, a water-cooled temperature-stabilized interrogation region and a high quality factor interrogation cavity. From the uncertainties of measured and calculated systematic frequency shifts, the fractional frequency accuracy is estimated to be 3.7 × 10−16. The fractional frequency stability, limited predominantly by noise in the local oscillator, is measured to be 7 × 10−16 after one day of averaging. Based on the proposed quasi-continuous regime of operation of the fountain, the accuracy of the Rb standard of 5 × 10−17 reachable in two days of averaging is predicted.