Abstract
This paper reports the measurement of the Neodymium isotopic composition by Neptune Multiple Collector Inductively Coupled Plasma Mass Spectrometry (MC-ICP-MS) over the last two years. Although there is concomitant Cerium in the chemical separation process, this has no significant influence on the Neodymium analysis. As for the sample containing small amounts of Samarium (Sm/Nd < 0.04), direct calibration for isobaric interference and mass discrimination by the exponential law can be obtained by assuming that Samarium mass discrimination is the same as that of Neodymium. Geological samples after traditional chemical separation were measured by Neptune MC-ICP-MS and Thermal Ionization Mass Spectrometry (TIMS) respectively. The results show that Neptune MC-ICP-MS can measure Neodymium isotopic composition as precisely the TIMS does and is even more effective and less time-consuming than the TIMS Method.