Abstract

Uniform guidelines for the data processing and validation of qualitative and quantitative multi-residue analysis using full-spectrum high-resolution mass spectrometry are scarce.Through systematic research, optimal mass accuracy and sensitivity are obtained after refining the post-processing of the HRMS data. For qualitative analysis, transforming the raw profile spectra to centroid spectra is recommended resulting in a 2.3 fold improved precision on the accurate mass determination of spectrum peaks. However, processing centroid data for quantitative purposes could lead to signal interruption when too narrow mass windows are applied for the construction of extracted ion chromatograms. Therefore, peak integration on the raw profile data is recommended. An optimal width of the mass window of 50ppm, which is a trade-off between sensitivity and selectivity, was obtained for a TOF instrument providing a resolving power of 20,000 at full width at half maximum (FWHM).For the validation of HRMS analytical methods, widespread concepts such as the signal-to-noise ratios for the determination of decision limits and detection capabilities have shown to be not always applicable because in some cases almost no noise can be detected anymore. A statistical methodology providing a reliable alternative is extended and applied.