Abstract

PurposeThe new computerized system is based on image analysis and designed to aid in orthopedic surgeries by virtual trajectory of the guide wire, intra-operative planning and various measurements. Validation of the accuracy and safety of any computer-aided surgery system is essential before implementing it clinically. We examined the accuracy of guide-wire length and angle measurements and fusion of multiple adjacent images (panoramic view image, PVI®) of the new software. MethodsThis is a 2-part study. Part I: twenty guide wires were drilled to various depths in a synthetic femur model and the results obtained by the software measurements were compared with manual measurements by a caliper and a depth gauge. Part II: a sawbone femur shaft was osteotomized and various inclinations of >10° to the varus or valgus angles were tested. The manually obtained measurements of angles and lengths were compared to the new computerized system software PVI. ResultsThere was a significant positive linear correlation between all groups of the computerized length and the control measurements (r>0.983, p<0.01). There was no significant difference among different distances, angles or positions from the image intensifier. There was a significant positive linear correlation between the angle and length measurement on the PVI and the control measurement (r>0.993, p<0.01). ConclusionsThe new computerized software has high reliability in performing measurements of length using an aiming, positioning and referring device intra-operatively.

Full Text
Published version (Free)