Abstract

This study examined how different methods of biometric measurement, types of intraocular lens calculation, and levels of surgeon experience affected predicted postoperative astigmatism with toric intraocular lens implantation. The error in predicted residual astigmatism was determined using the IOL Master 500 and Lenstar LS 900 devices and was calculated by the Alcon toric calculator, Alcon toric calculator with Baylor nomogram, and Barrett toric calculator. Results were reviewed from 147 eyes of 125 patients. Of the three calculations, the Barrett toric calculator had the lowest mean absolute error in predicted residual astigmatism (P < 0.0001). The mean absolute errors of the IOL Master 500 and the Lenstar LS 900 were not significantly different when compared using three different calculation methods (P = 0.54). The mean absolute error was significantly lower (P = 0.049) for ophthalmology staff than for ophthalmology residents when calculation methods were considered. In conclusion, the calculation method used for IOL selection in cataract surgery significantly affects the accuracy of both staff and residents.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.