Abstract

Introduction The aim of this study was to determine in anterior teeth, bicuspids, and molars (1) the accuracy of 3 different electronic apex locators (EALs) in detecting the apical foramen and (2) the accuracy of digital radiography in determining the working length (WL), compared with visible control under a microscope. Methods By using radiovideography (RVG), we measured the lengths of 120 root canals with 3 different EALs (Endex, ProPex II, and Root ZX) and compared them with the actual lengths. The accuracy of EALs and RVG was related to each dental category. An endodontic training kit (Pro-Train) was used during experimental procedures. Results Statistical analysis showed that the 3 EALs and RVG were less accurate in anterior teeth and molars than in bicuspids. The paired-sample t test showed no statistically significant difference between mesiodistal plane and buccolingual plane digital radiography in all groups. Conclusions The 3 EALs tested were more accurate in detecting the apical foramen in bicuspids than in both molars and anterior teeth. Radiographic measurements were not reliable for determining WL in all dental groups in both radiographic planes.