Abstract
The purpose of this study was to evaluate the accuracy of a method for estimating the tip position of a fiber optic shape-sensing (FOSS) integrated instrument being inserted through a bronchoscope. A modified guidewire with a multicore optical fiber was inserted into the working channel of a custom-made catheter with three electromagnetic (EM) sensors. The displacement between the instruments was manually set, and a point-based method was applied to match the position of the EM sensors to corresponding points on the shape. The accuracy was evaluated in a realistic bronchial model. An additional EM sensor was used to sample the tip of the guidewire, and the absolute deviation between this position and the estimated tip position was calculated. For small displacements between the tip of the FOSS integrated tool and the catheter, the median deviation in estimated tip position was ≤5 mm. For larger displacements, deviations exceeding 10 mm were observed. The deviations increased when the shape sensor had sharp curvatures relative to more straight shapes. The method works well for clinically relevant displacements of a biopsy tool from the bronchoscope tip, and when the path to the lesion has limited curvatures. However, improvements must be made to our configuration before pursuing further clinical testing.