Abstract

ABSTRACTObjectives:Wearable devices such as fitness trackers have become popular in the healthcare field. Tracking heart rate and respiratory rate, in addition to physical activity, may provide an accurate picture of daily health. We believe that a combination of two types of devices can simultaneously measure and record physical activity, heart rate, and respiratory rate. However, the measurement accuracies of these two types of devices are not clear. This study aimed to determine the measurement accuracies of two wearable devices for heart and respiratory rate measurements.Methods:Ten healthy men performed incremental load tests (ILTs) and constant load tests (CLTs) on a cycle ergometer. The heart and respiratory rates were measured using wrist-worn (Silmee W22, TDK, Japan, Tokyo) and respiratory tracking devices (Spire Stone, Spire Health, San Francisco, CA, USA), respectively. A 12-lead electrocardiograph and the breath-by-breath method were used as external standards for heart and respiratory rates, respectively.Results:Bland–Altman analysis showed that heart rate had a fixed bias at rest and during ILT and CLT and had a proportional bias during CLT. The standard error values of the regression at rest and during CLT were less than 10 bpm for heart rate and less than 5.0 /min for respiratory rate. During ILT, the standard error was greater than 10 bpm for heart rate and approximately 5.0 /min for respiratory rate.Conclusions:The heart and respiratory rate measurements obtained using wearable devices were accurate within the practical margin of error.