Abstract
<b>Background: </b>Advances in continuous glucose monitoring (CGM) have transformed ambulatory diabetes management. Until recently, inpatient use of CGM has remained investigational with limited data on its accuracy in the hospital setting. <p><b>Methods: </b>To analyze the accuracy of Dexcom G6 CGM,<b> </b>we compared retrospective matched-pair CGM and capillary point-of-care (POC) glucose data from three inpatient CGM studies (two interventional and one observational) in general medicine and surgery patients with diabetes treated with insulin. Analysis of accuracy metrics included mean absolute relative difference (MARD), median absolute relative difference (ARD), and proportion of CGM values within ±15, 20 and 30% or ±15, 20 and 30 mg/dL of POC reference values for blood glucose >100 mg/dL or ≤100 mg/dL, respectively (?/15, /20, 0/30). Clinical reliability was assessed using Clarke error grid analyses.</p> <p><b>Results: </b>A total of 218 patients were included (96% with type 2 diabetes) with a mean age of 60.6 ± 12 years. The overall MARD (n=4,067 matched glucose pairs) was 12.8% and median ARD was 10.1% [IQR 4.6, 17.6]. The proportion of readings meeting ?/15, /20 and 0/30 criteria were 68.7, 81.7, and 93.8%. Clarke error grid analysis showed 98.7% of all values in zones A+B. MARD and median ARD were higher in hypoglycemia (<70mg/dL) and severe anemia (hemoglobin <7g/dL).</p> <p><b>Conclusion: </b>Our results indicate that CGM technology is a reliable tool for hospital use and may help improve glucose monitoring in non-critically ill hospitalized patients with diabetes. </p>