Abstract

PurposeTo compare and analyze the interchangeability of three measuring systems, each based on a different technique, for central corneal thickness (CCT) analysis. MethodsCCT measurements were measured using optical coherence tomography (OCT), non-contact specular microscopy (NCSM), and ultrasonic pachymetry (USP) in 60 eyes of 60 healthy patients with a mean age of 66.5±15.0 years and a mean spherical equivalent of 0.43±1.14 D. Analysis of variations in measurement concordance and correlation among the three different methods were performed. Comparison of CCT measurements were done using Bland–Altman plots (with bias and 95% confidence intervals), intraclass correlation coefficient (ICC), and paired t-student analysis. ResultsMean CCT values were: 549.20±26.91μm for USP (range 503–618μm), 514.20±27.49μm for NCSM (range 456–586μm) and 542.80±25.56μm for OCT (range 486–605μm). CCT values obtained with NCMS were significantly lower than those obtained with OCT and USP methods. NCMS CCT value was 36.08±10.72μm lower than USP value (p<0.05), and NCMS CCT value was 7.88±8.86μm lower than OCT value (p<0.05). ICC between USP–NCSM pair was 0.488 and 0.909 between USP–OCT pair. ConclusionOCT and UPS offered highly comparable results, whereas NCSM offered lower mean CCT values compared to the other two methods. Therefore, NCSM should not be considered a reliable method for measuring CCT and should rather be considered for assessing longitudinal changes in the same patient.