Abstract

The validity of the Hanau ear-bow to transfer an arbitrary hinge axis to a Hanau articulator was clinically compared with a Hanau kinematic face-bow. The study was conducted with 18 randomly selected patients. This investigation demonstrated a significant statistical difference between the arbitrary axis located with an ear-bow and the terminal hinge axis. This discrepancy was significant in the anteroposterior direction but not in the superior-inferior direction. Only 50% of the arbitrary hinge axes were within a 5 mm radius of the terminal hinge axis, while 89% were within a 6 mm radius. Furthermore, the ear-bow method was not repeatable statistically. Additional study is needed to determine the practical value of the arbitrary face-bow and to pursue modifications to improve its accuracy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.