Abstract

This study determined the effect on articulator settings of using an arbitrary hinge axis as compared with the terminal hinge axis in orienting a computerized pantograph on a patient. The effect of an error in measuring the reference plane angle or clutch angle was also determined. Two subjects were pantographed 30 times by using the terminal hinge axis and 30 times by using the arbitrary hinge axis. Statistically the articulator settings obtained by using the arbitrary hinge axis and the terminal hinge axis were different, but the difference was small. One subject was pantographed on 18 occasions by using different reference plane angle and clutch angle settings. Analysis showed that 1-degree change in the reference plane angle setting produced ~0.9-degree change in the protrusive and orbiting path articulator settings. A 1-degree change in the clutch angle setting produced ~0.9-degree change in the progressive side shift settings. It was concluded that the terminal hinge axis should be used as the posterior reference point and the reference plane angle and clutch angle should be carefully recorded.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.