Abstract

The accuracy of a portable blood gas analyzer (OPTI 1) was evaluated using canine blood and aqueous control solutions. Sixty-four arterial blood samples were collected from 11 anesthetized dogs and were analyzed for pH, partial pressure of carbon dioxide (PCO2) partial pressure of oxygen (PO2), and bicarbonate concentration ([HCO3-]) values by the OPTI 1 and a conventional blood gas analyzer (GASTAT 3). The conventional analyzer was considered as a standard against which the OPTI 1 was evaluated. Comparison of OPTI 1 results with those of GASTAT 3 by linear regression analysis revealed a high degree of correlation with the GASTAT 3 (r = .90-.91). The mean +/- SD of the differences between OPTI 1 and GASTAT 3 values was -0.008 +/- 0.017 for pH, -0.88 +/- 3.33 mm Hg for PCO2, 3.71 +/- 6.98 mm Hg for PO2, and -0.34 +/- 1.45 mEq/L for [HCO3-]. No statistically significant difference was found between the OPTI 1 and the GASTAT 3. Agreement between these 2 methods is within clinically acceptable ranges for pH, PCO2, PO2, and [HCO3-]. The coefficients of variation for measured pH, PCO2, and PO2 values of 3 aqueous control solutions (acidic, normal, and alkalotic) analyzed by the OPTI 1 ranged from 0.047 to 0.072% for pH, 0.78 to 1.81% for PCO2, and 0.73 to 2.77% for PO2. The OPTI 1 is concluded to provide canine blood gas analysis with an accuracy that is comparable with that of conventional benchtop blood gas analyzers.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.