Abstract

PurposeHigh tibial osteotomy (HTO) is an effective treatment option in early osteoarthritis. However, preoperative planning and surgical execution can be challenging. Computer assisted three-dimensional (3D) planning and patient-specific instruments (PSI) might be helpful tools in achieving successful outcomes. Goal of this study was to assess the accuracy of HTO using PSI.MethodsAll medial open wedge PSI-HTO between 2014 and 2016 were reviewed. Using pre- and postoperative radiographs, hip-knee-ankle angle (HKA) and posterior tibial slope (PTS) were determined two-dimensionally (2D) to calculate 2D accuracy. Using postoperative CT-data, 3D surface models of the tibias were reconstructed and superimposed with the planning to calculate 3D accuracy.ResultsTwenty-three patients could be included. A mean correction of HKA of 9.7° ± 2.6° was planned. Postoperative assessment of HKA correction showed a mean correction of 8.9° ± 3.2°, resulting in a 2D accuracy for HKA correction of 0.8° ± 1.5°. The postoperative PTS changed by 1.7° ± 2.2°. 3D accuracy showed average 3D rotational differences of − 0.1° ± 2.3° in coronal plane, − 0.2° ± 2.3° in transversal plane, and 1.3° ± 2.1° in sagittal plane, whereby 3D translational differences were calculated as 0.1 mm ± 1.3 mm in coronal plane, − 0.1 ± 0.6 mm in transversal plane, and − 0.1 ± 0.6 mm in sagittal plane.ConclusionThe use of PSI in HTO results in accurate correction of mechanical leg axis. In contrast to the known problem of unintended PTS changes in conventional HTO, just slight changes of PTS could be observed using PSI. The use of PSI in HTO might be preferable to obtain desired correction of HKA and to maintain PTS.

Highlights

  • High tibial osteotomy (HTO) is an effective treatment option in early osteoarthritis

  • Postoperative assessment of hip-knee-ankle angle (HKA) correction showed a mean correction of 8.9° ± 3.2°, resulting in a 2D accuracy for HKA correction of 0.8° ± 1.5°

  • The postoperative posterior tibial slope (PTS) changed by 1.7° ± 2.2°. 3D accuracy showed average 3D rotational differences of − 0.1° ± 2.3° in coronal plane, − 0.2° ± 2.3° in transversal plane, and 1.3° ± 2.1° in sagittal plane, whereby 3D translational differences were calculated as 0.1 mm ± 1.3 mm in coronal plane, − 0.1 ± 0.6 mm in transversal plane, and − 0.1 ± 0.6 mm in sagittal plane

Read more

Summary

Methods

All medial open wedge PSI-HTO between 2014 and 2016 were reviewed. Using postoperative CT-data, 3D surface models of the tibias were reconstructed and superimposed with the planning to calculate 3D accuracy. Patient selection All patients who underwent medial open wedge HTO using PSI from March 2014 until August 2016 were included. Indication for HTO was set in patients with medial varus osteoarthritis Grade 1–3 according to the Kellgren and Lawrence classification [15], with an age under 60 years and with failed conservative treatment. Cases performed without PSI or in which the surgeon intraoperatively intentionally modified the preoperative plan, were excluded.

Results
Discussion
Conclusion