Abstract

In this work we present the 2 MV Tandetron accelerator manufactured by High Voltage Engineering Europa (HVEE), that was installed at MTA Atomki in Debrecen, Hungary in 2015. Furthermore, we report on the first performance test of the whole facility including the calibration of the terminal voltage using accurately known resonance energies of the 27Al(p,γ)28Si and 13C(p,γ)14N reactions, as well as neutron threshold energies of the 7Li(p,n)7Be and 13C(p,n)13N reactions. The accurate energy calibration of the generating voltmeter (GVM), the good energy stability of the accelerator as well as the low energy spread of the ion beam suited the requirements of nuclear astrophysics, resonance PIGE depth profiling and the nanoprobe beamline. We have investigated whether the terminal voltage reading depends on the insulating gas pressure, and how much the stripper gas pressure modifies the beam energy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.