Abstract

Prostate T(2) mapping was performed in 34 consecutive patients using an accelerated multiecho spin-echo sequence with 4-fold k-space undersampling leading to a net acceleration factor of 3.3 on a 3T scanner. The mean T(2) values from the accelerated and conventional, unaccelerated sequences demonstrated a very high correlation (r = 0.99). Different prostate segments demonstrated similarly good interscan reproducibility (p = not significant) with slightly larger difference at base: 2.0% ± 1.6% for left base and 2.1% ± 1.1% for right base. In patients with subsequent targeted biopsy, T(2) values of histologically proven malignant tumor areas were significantly lower than the suspicious looking but nonmalignant lesions (p < 0.05) and normal areas (p < 0.001): 100 ± 10 ms for malignant tumors, 114 ± 23 ms for suspicious lesions and 149 ± 32 ms for normal tissues. The proposed method can provide an effective approach for accelerated T(2) quantification for prostate patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.