Abstract
Large magnetic field inhomogeneity can be a significant cause of spatial flip-angle variation when using ordinary, limited-bandwidth RF pulses. Multidimensional RF pulses are particularly sensitive to inhomogeneity due to their extended pulse length, which decreases their bandwidth. Previously, it was shown that, by breaking a 2D pulse into multiple undersampled k-space segments, the excitation bandwidth can be increased at the expense of increased imaging time. The present study shows how this increased imaging time can be offset by undersampling acquisition k-space in a phase-encoded dimension that is in the direction of excitation segmentation. Data from each segment are viewed as originating from "virtual receive coils" rather than multiple physical coils. The undersampled data are reconstructed using parallel imaging techniques (e.g. as in GRAPPA). The method was tested in vivo with brain imaging at both 3 T and 4 T, and used in conjunction with a 32-channel head coil and conventional GRAPPA on the 3 T data. Relationships with existing techniques and future applications are discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of magnetic resonance (San Diego, Calif. : 1997)
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.